Abstract
Previous studies have not demonstrated the usefulness of a mechanical ventilator with the Universal Portable Anesthesia Complete (UPAC) field anesthesia delivery system in a pushover mode. This study demonstrated that the Lifecare PLV-100 ventilator can function effectively in a practical pushover configuration with the UPAC vaporizer. By comparison, vaporizer output followed the patterns of documented concentration curves for isoflurane at a given dial setting and minute ventilation. Measured airway pressures in the breathing circuit were within physiological parameters.
Original language | English (US) |
---|---|
Pages (from-to) | 164-168 |
Number of pages | 5 |
Journal | Military medicine |
Volume | 163 |
Issue number | 3 |
DOIs | |
State | Published - Mar 1998 |
Externally published | Yes |
ASJC Scopus subject areas
- Public Health, Environmental and Occupational Health