N2 - Aim. The purpose of this study was to evaluate in vitro differences of colony forming units (CFUs), of four different bacteria (Staphylococcus epidermidis, Staphylococcus aureus, Pseudomonas aeruginosa, Escherichia coli), over 4days, using bovine blood, in five intravenous needleless connectors (three negative, one positive, one zero). Background. Availability of colony forming units creates a definitive environment for bloodstream infections and occlusions. Protection of the intraluminal pathway is one important way to help eliminate occlusions and catheter-related bloodstream infections and utilization of best available product(s) will aid in best patient outcomes. Methods. Five different connectors evaluated in 2009 by an independent laboratory for in vitro differences about colony forming units of four different organisms over 4days. Results. The Q-Syte™ performed poorly on all organisms (P<0·0001). The MaxPlus® Clear and MicroCLAVE® fluctuated between high colony forming units (28·15 & 56·55 respectively) and zero colony forming units. The TKO™Clave® stayed increased (high of 50·8 colony forming units). The RyMed-5001® performed the best with very low colony forming units (2·25 CFUs to zero). Conclusions. Non-antimicrobial connectors differ on colony forming unit counts in vitro for four types of bacteria. Connectors with most colony forming units to least colony forming units included the Q-Syte™, TKO™Clave®, MicroCLAVE®, MaxPlus® Clear and RyMed-5001®. Connectors are one statistically significant variable (50%) in the development of occlusions and infections. Staff nurses, managers, infection control specialists and vascular access specialists in all settings need to use technologies that invoke the least patient harm. The RyMed-5001® connector best protects the intraluminal pathway from bacteria compared with four other commonly used connectors in vitro.
AB - Aim. The purpose of this study was to evaluate in vitro differences of colony forming units (CFUs), of four different bacteria (Staphylococcus epidermidis, Staphylococcus aureus, Pseudomonas aeruginosa, Escherichia coli), over 4days, using bovine blood, in five intravenous needleless connectors (three negative, one positive, one zero). Background. Availability of colony forming units creates a definitive environment for bloodstream infections and occlusions. Protection of the intraluminal pathway is one important way to help eliminate occlusions and catheter-related bloodstream infections and utilization of best available product(s) will aid in best patient outcomes. Methods. Five different connectors evaluated in 2009 by an independent laboratory for in vitro differences about colony forming units of four different organisms over 4days. Results. The Q-Syte™ performed poorly on all organisms (P<0·0001). The MaxPlus® Clear and MicroCLAVE® fluctuated between high colony forming units (28·15 & 56·55 respectively) and zero colony forming units. The TKO™Clave® stayed increased (high of 50·8 colony forming units). The RyMed-5001® performed the best with very low colony forming units (2·25 CFUs to zero). Conclusions. Non-antimicrobial connectors differ on colony forming unit counts in vitro for four types of bacteria. Connectors with most colony forming units to least colony forming units included the Q-Syte™, TKO™Clave®, MicroCLAVE®, MaxPlus® Clear and RyMed-5001®. Connectors are one statistically significant variable (50%) in the development of occlusions and infections. Staff nurses, managers, infection control specialists and vascular access specialists in all settings need to use technologies that invoke the least patient harm. The RyMed-5001® connector best protects the intraluminal pathway from bacteria compared with four other commonly used connectors in vitro.
KW - Bloodstream
KW - Infection
KW - Intravenous connector
KW - Intravenous technology
UR - http://www.scopus.com/inward/record.url?scp=79958755048&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=79958755048&partnerID=8YFLogxK
U2 - 10.1111/j.1365-2648.2010.05598.x
DO - 10.1111/j.1365-2648.2010.05598.x
M3 - Article
C2 - 21366670
AN - SCOPUS:79958755048
SN - 0309-2402
VL - 67
SP - 1601
EP - 1613
JO - Journal of Advanced Nursing
JF - Journal of Advanced Nursing
IS - 7
ER -