Abstract
A sensitive high-performance chromatographic method for the detection of amino acids and oligopeptides with pre-column fluorescence derivatization has been developed. Glycine, glycylglycine, triglycine, glutathion, glutamic acid, and cysteine were separated on a reversed-phase C18 column with methanol-water-triethylamine as eluent; the derivatization and the chromatographic conditions were optimized. The six derivatives were eluted in 20 rain with good reproducibility. The relative standard deviations (n=6) at an analytical concentration of 2 x 10-6 M are less than 4.5%. The detection limits (signal-to-noise ratio=3) for the six derivatives are in the range of 15-46 fmol.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 2103-2115 |
| Number of pages | 13 |
| Journal | Journal of Liquid Chromatography and Related Technologies |
| Volume | 21 |
| Issue number | 14 |
| DOIs | |
| State | Published - 1998 |
| Externally published | Yes |
ASJC Scopus subject areas
- Analytical Chemistry
- Biochemistry
- Pharmaceutical Science
- Clinical Biochemistry
Fingerprint
Dive into the research topics of 'High-performance liquid chromatographic determination of amino acids and oligopeptides by pre-column fluorescence derivatization with 9-fluorenyl- methoxy carbonyl succinimide'. Together they form a unique fingerprint.Cite this