Abstract
Stereotactic radiosurgery has become a widely used technique to treat solid tumors and secondary metastases of the brain. Multiple targets can be simultaneously treated with a single isocenter in order to reduce the set-up time to improve patient comfort and workflow. In this study, a 5-arc multifocal RapidArc treatment was delivered to multiple PRESAGE® dosimeters in order to explore the repeatability of the treatment. The three delivery measurements agreed well with each other, with less than 3% standard deviation of dose in the target. The deliveries also agreed well with the treatment plan, with gamma passing rates greater than 90% (5% dose-difference, and 2 mm distance-to-agreement criteria). The optical-CT PRESAGE® system provided a reproducible measurement for treatment verification, provided measurements were made immediately following treatment.
Original language | English (US) |
---|---|
Article number | 012072 |
Journal | Journal of Physics: Conference Series |
Volume | 444 |
Issue number | 1 |
DOIs | |
State | Published - 2013 |
Externally published | Yes |
Event | 7th International Conference on 3D Radiation Dosimetry, IC3DDose 2012 - Sydney, NSW, Australia Duration: Dec 4 2012 → Dec 8 2012 |
ASJC Scopus subject areas
- Physics and Astronomy(all)