Abstract
Platinum microelectrodes modified with a lipid bilayer membrane incorporating cholesterol oxidase are used for detection of cholesterol contained in the plasma membrane of a single cell. Amperometric responses are consistent with enzymatic catalysis being rate limiting and cholesterol diffusing laterally in the plasma membrane to the electrode contact site. Importantly, electrode response appears to correlate with the cholesterol content of the cell plasma membrane. The electrodes should be useful for characterizing cellular cholesterol tracking pathways involved in pathogenesis of disease.
Original language | English (US) |
---|---|
Pages (from-to) | 10214-10215 |
Number of pages | 2 |
Journal | Journal of the American Chemical Society |
Volume | 126 |
Issue number | 33 |
DOIs | |
State | Published - Aug 25 2004 |
ASJC Scopus subject areas
- Catalysis
- Chemistry(all)
- Biochemistry
- Colloid and Surface Chemistry