Skip to search boxSkip to navigationSkip to main content

Benchtop testing and comparisons among three types of through-the-scope endoscopic clipping devices

  • Sumanth R. Daram
    ,
  • Shou Jiang Tang(corresponding author)
    ,
  • Ruonan Wu
    ,
  • S. D.Filip To
*Corresponding author for this work
  • University of Mississippi
    ,
  • Mississippi State University
Scholary Output:
Contribution to journal
Article
Peer-review

Abstract

Background: Through-the-scope (TTS) endoscopic clipping devices are widely used. No benchtop testing or direct comparisons of these endoclips have been performed to show their rotational ability and inherent mechanical strengths during closure and after deployment. This study aimed to provide benchtop data that can be used to guide clinical applications and to promote future device research and development. Methods: Benchtop testing and comparisons were performed for three groups of TTS clips: QuickClip2 long, resolution, and instinct clips. The main outcome measurements were device-in-endoscope retroflection angles (DIERA), opening strength, "snapping" force of acute clip closure, and neoprene pulling strength. Results: The achievable gastroscope DIERA was 10 for QuickClip2, 3 for the resolution clip, and 10 for the instinct clip. The QuickClip and the Instinct clip rotated almost equally well under all endoscope configurations, including endoscopic retrograde cholangiopancreatography (ERCP). With or without a sheath, the resolution clip lacked the ability to rotate. During clip opening force testing (the amount of force required to force open the jaws of a deployed clip by 3.2 mm; 3.2 mm was chosen due to the standard dimension of the gauge used for the measurement), the Instinct clips were the strongest. For the Instinct clips, an opening force of 404 ± 124 g was needed to open the closed clip, and an additional 386 ± 133 g was required to open the clip jaws to 3.2 mm. In terms of snapping force during acute closure and neoprene pulling strength, the instinct and resolution clips performed almost equally. The limitations of the study were the benchtop testing and the finite sample size for closing and pulling strength comparisons. Conclusions: The QuickClip2 and the Instinct clip rotate equally well under different endoscope configurations. The resolution clips lack rotational ability. The instinct clips are stronger mechanically than the other two TTS clips. Stronger clips are perhaps associated with higher therapeutic efficacy and retention rates.

Publication Information

Output type

Scholary Output:
Contribution to journal
Article
Peer-review

Original language

English (US)

Pages from-to (Number of pages)

Pages 1521-1529 (9 pages)

Journal (Volume, Issue Number)

Surgical endoscopy (Volume 27, Issue 5)

Publication milestones

  • Published - 05/2013

Publication status

Published - 05/2013

ISSN

0930-2794

Publication IDs

  • Scopus: 84876281438
  • PubMed: 23292554

Publication metrics

Metrics

Scopus
citations
SciVal
citations
24
Fractional count
1
Fractional count
0.25
Fractional count
3
Fractional count
0.75
Fractional count
1
Fractional count
1
SciVal
FWCI
2.42
SciVal
Author count
4
SciVal
Paper percentile
83

PlumX, opens in new tab

Citation count
30
Captures
21