Skip to search boxSkip to navigationSkip to main content

Bioprosthetic valve fracture: Technical insights from a multicenter study

  • Bioprosthetic Valve Fracture Investigators
  • ,
  • Georgetown University
    ,
  • Intermountain Heart Institute
    ,
  • Morton Plant Mease Health Care
    ,
  • Saint Louis University
    ,
  • University of British Columbia
Scholary Output:
Contribution to journal
Article
Research
Peer-review

Open access

Abstract

Objective: Valve-in-valve transcatheter aortic valve replacement (VIV TAVR) can result in high residual gradients that are associated with increased mortality. Bioprosthetic valve fracture (BVF) has been shown to improve residual gradients following VIV TAVR; however, factors influencing the results of BVF have not been studied. Methods: BVF was performed in 75 patients at 21 centers. Hierarchical multiple linear regression was performed to identify variables that were associated with lower final transvalvular gradient. Results: Surgical valves with a median true internal diameter of 18.5 mm (interquartile range, 17.0-20.5 mm) were treated with VIV TAVR in conjunction with BVF using balloon-expandable (n = 43) or self-expanding (n = 32) transcatheter heart valves with a median size of 23 mm (interquartile range, 23-23 mm). There were no aortic root disruptions, coronary occlusions, or new pacemakers; in-hospital or 30-day mortality was 2.6% (2 out of 75). Final mean transvalvular gradient was 9.2 ± 6.3 mm Hg, but was significantly lower when BVF was performed after VIV TAVR compared with BVF first (8.1 ± 4.8 mm Hg vs 16.9 ± 10.1 mm Hg; P <.001). After adjusting for timing of BVF (ie, before or after VIV TAVR), transcatheter heart valve size/type, surgical valve mode of failure, true internal diameter, and baseline gradient and BVF balloon size, performing BVF after VIV TAVR (P <.001) and using a larger BVF balloon (P =.038) were the only independent predictors of lower final mean gradient. Conclusions: BVF can be performed safely and results in reduced residual transvalvular gradients. Performing BVF after VIV TAVR and using larger balloon appears to achieve the best hemodynamic results.

Publication Information

Output type

Scholary Output:
Contribution to journal
Article
Research
Peer-review

Original language

English (US)

Pages from-to (Number of pages)

Pages 1317-1328.e1

Journal (Volume, Issue Number)

Journal of Thoracic and Cardiovascular Surgery (Volume 158, Issue 5)

Publication milestones

  • Published - 11/2019

Publication status

Published - 11/2019

ISSN

0022-5223

Publication IDs

  • Scopus: 85062405632
  • PubMed: 30857820

Publication metrics

Metrics

Fractional count
1
Fractional count
0.02
Fractional count
44
Fractional count
0.98
Fractional count
1
Fractional count
1
SciVal
citations
35
SciVal
FWCI
9.12
SciVal
Author count
45
SciVal
Paper percentile
98
SciVal
Top percentile
5
Scopus
citations

PlumX, opens in new tab

Citation count
101
Captures
62

Funding Details

Drs Allen and Chhatriwalla have performed research/clinical trial support/proctoring/speakers bureau duties for Abbott Vascular, Edwards Lifesciences, and Medtronic. Dr Cohen has performed research/clinical trial support for Medtronic, Abbott Vascular, Edwards Lifesciences, and Boston Scientific as well as consulting services for Medtronic and Edward Lifesciences. Dr Rovin has been a proctor/consultant for Medtronic and a proctor and speakers bureau member for Abbott. Dr Nguyen has performed research/clinical trial support/proctoring for Abbott Vascular, Edwards Lifesciences, and LivaNova. Dr Dvir has provided research/clinical trial support/consulting for Medtronic, Abbott Vascular, and Edwards Lifesciences. Dr Webb has been a consultant for Edwards Lifesciences and Abbott Vascular and has performed research/clinical trial support/proctoring/speakers bureau duties for Medtronic, Abbott Vascular, and Edwards Lifesciences. Dr Whisenant has provided consulting services for Boston Scientific and Edwards Lifesciences. Dr Bavry has received travel support from Edwards Lifesciences.
FundersFunding numbers
Boston Scientific and Edwards Lifesciences
-
Medtronic
-
Edwards Lifesciences
-
BSC
-
Abbott Vascular
-
LivaNova PLC
-