Skip to search boxSkip to navigationSkip to main content

CALAXO osteoconductive interference screw: the value of postmarket surveillance.

  • Charles L. Cox(corresponding author)
    ,
  • Kelly C. Homlar
    ,
  • James L. Carey
    ,
  • Kurt P. Spindler
*Corresponding author for this work
  • Vanderbilt University
    ,
  • Unknown
Scholary Output:
Contribution to journal
Article
Research
Peer-review

Abstract

The CALAXO osteoconductive interference screw was recalled in August 2007 due to reports of increased numbers of postoperative complications associated with screw swelling and prominence leading to the need for surgical debridement. This study reviews complications associated with CALAXO screw use in a consecutive cohort of patients undergoing anterior cruciate ligament reconstruction surgery by the senior author at the authors' institution. Over a 12-month period, 226 CALAXO interference screws, either of 20 mm length or 25 mm length, were implanted in 112 patients, and postoperative complications were noted. The 25-mm tibial screw was over 5 times (RR 5.2, 95% CI 1.8 to 15.3) more likely to be prominent than the 20-mm screw (p value=.002). Four surgical debridements were required in the 25-mm tibial screw group; none were required in the 20-mm group. The authors hypothesize that the inability to bury the longer screw length into the bone tunnel is associated with postoperative complications associated with the CALAXO screw.

Publication Information

Output type

Scholary Output:
Contribution to journal
Article
Research
Peer-review

Original language

English (US)

Pages from-to (Number of pages)

Pages 121-124 (4 pages)

Journal (Volume, Issue Number)

Journal of Surgical Orthopaedic Advances (Volume 19, Issue 2)

Publication milestones

  • Published - 2010

Publication status

Published - 2010

ISSN

1548-825X

Publication IDs

  • Scopus: 77957010239
  • PubMed: 20727309

Publication metrics

Metrics

Scopus
citations
SciVal
FWCI
0.26
SciVal
Author count
4
SciVal
citations
7
SciVal
Paper percentile
57
Fractional count
1
Fractional count
0.25
Fractional count
3
Fractional count
0.75
Fractional count
1
Fractional count
1

PlumX

Captures
18
Citation count
11

Funding Details

FunderFunding number
NIAMS
R01AR053684