Cardiac cine CT approaching 1 mSv: implementation and assessment of a 58-ms temporal resolution protocol

  • Younhee J. Choi
    ,
  • Mark A. Ahlman
    ,
  • Marissa Mallek
    ,
  • Tyler E. Cork
    ,
  • Marcus Y. Chen
    ,
  • David A. Bluemke
*Corresponding author for this work
  • National Institutes of Health, Bethesda
    ,
  • National Institutes of Health
    ,
  • University of Wisconsin-Madison
    ,
  • Stanford University
Scholary Output:
Contribution to journal
Article
Research
Peer-review

Abstract

Clinical use of cardiac cine CT imaging is limited by high radiation dose and low temporal resolution. To evaluate a low radiation dose, high temporal resolution cardiac cine CT protocol in human cardiac CT and phantom scans. CT scans of a circulating iodine target were reconstructed using the conventional single heartbeat half-scan (HS, approx. 175 ms temporal resolution) and the 3-heartbeat multi-segment (MS, approx. 58 ms) algorithms. Motion artifacts were quantified by the root-mean-square error (RMSE). Low-dose cardiac cine CT scans were performed in 55 subjects at a tube potential of 80 kVp and current of 80 mA. Image quality of HS and MS scans was assessed by blinded reader quality assessment, left ventricular (LV) free wall motion, and LV ejection rate. Motion artifacts in phantom scans were higher in HS than in MS reconstructions (RSME 188 and 117 HU, respectively; p = 0.001). Median radiation dose in human scans was 1.2 mSv. LV late diastolic filling was observed more frequently in MS than in HS images (42 vs. 26 subjects, respectively; p < 0.001). LV free wall systolic motion was more physiologic and had less error in MS than in HS reconstructions (sum-of-squared errors 34 vs. 45 mm2, respectively; p < 0.001), and LV peak ejection rate was higher in MS than in HS reconstructions (166 vs. 152 mL/s, respectively; p < 0.001). Cardiac cine CT imaging is feasible at a low radiation dose of 1.2 mSv. MS reconstruction showed improved imaging of rapid motion in phantom studies and human cardiac CTs.

Publication Information

Output type

Scholary Output:
Contribution to journal
Article
Research
Peer-review

Original language

English (US)

Pages from-to (Number of pages)

Pages 1583-1591 (9 pages)

Journal (Volume, Issue Number)

International Journal of Cardiovascular Imaging (Volume 36, Issue 8)

Publication milestones

  • Published - 08/01/2020

Publication status

Published - 08/01/2020

ISSN

1569-5794

Publication IDs

  • Scopus: 85085092937
  • PubMed: 32367189

Publication metrics

Metrics

Fractional count
1
Fractional count
0.14
Fractional count
6
Fractional count
0.86
Fractional count
1
Fractional count
1
Scopus
citations

PlumX, opens in new tab

Captures
12
Citation count
5

Funding Details

We thank Rolf Symons, currently at the University Hospitals Leuven, for his constructive feedback on the manuscript and Amir Pourmorteza, currently at Emory University, for his critical review of the manuscript.
FundersFunding number
University Hospitals Leuven
-
NHLBI
ZIAHL006220