Commissioning a small-field biological irradiator using point, 2D, and 3D dosimetry techniques

  • ,
  • Mark Oldham(corresponding author)
    ,
  • Andrew Thomas
    ,
  • Yifan Li
    ,
  • John Adamovics
    ,
  • David G. Kirsch
*Corresponding author for this work
  • Duke University
    ,
  • Rider University
Scholary Output:
Contribution to journal
Article
Peer-review

Open access

Abstract

Purpose: To commission a small-field biological irradiator, the XRad225Cx from Precision x-Ray, Inc., for research use. The system produces a 225 kVp x-ray beam and is equipped with collimating cones that produce both square and circular radiation fields ranging in size from 1 to 40 mm. This work incorporates point, 2D, and 3D measurements to determine output factors (OF), percent-depth-dose (PDD) and dose profiles at multiple depths. Methods: Three independent dosimetry systems were used: ion-chambers (a farmer chamber and a micro-ionisation chamber), 2D EBT2 radiochromic film, and a novel 3D dosimetry system (DLOSPRESAGE). Reference point dose rates and output factors were determined from in-air ionization chamber measurements for fields down to ∼13 mm using the formalism of TG61. PDD, profiles, and output factors at three separate depths (0, 0.5, and 2 cm), were determined for all field sizes from EBT2 film measurements in solid water. Several film PDD curves required a scaling correction, reflecting the challenge of accurate film alignment in very small fields. PDDs, profiles, and output factors were also determined with the 3D DLOSPRESAGE system which generated isotropic 0.2 mm data, in scan times of 20 min. Results: Surface output factors determined by ion-chamber were observed to gradually drop by ∼9 when the field size was reduced from 40 to 13 mm. More dramatic drops were observed for the smallest fields as determined by EBT∼18 and ∼42 for the 2.5 mm and 1 mm fields, respectively. PRESAGE and film output factors agreed well for fields 20 mm (where 3D data were available) with mean deviation of 2.2 (range 1-4). PDD values at 2 cm depth varied from ∼72 for the 40 mm field, down to ∼55 for the 1 mm field. EBT and PRESAGE PDDs agreed within ∼3 in the typical therapy region (1-4 cm). At deeper depths the EBT curves were slightly steeper (2.5 at 5 cm). These results indicate good overall consistency between ion-chamber, EBT2 and PRESAGE measured OFs, PDDs, and profiles. Conclusions: The combination of independent 2D and 3D measurements was found to be valuable to ensure accurate and comprehensive commissioning. Film measurements were time consuming and challenging due to the difficulty of film alignment in small fields. PRESAGE 3D measurements were comprehensive and efficient, because alignment errors are negligible, and all parameters for multiple fields could be obtained from a single dosimeter and scan. However, achieving accurate superficial data (within 4 mm) is not yet feasible due to optical surface artifacts.

Publication Information

Output type

Scholary Output:
Contribution to journal
Article
Peer-review

Original language

English (US)

Pages from-to (Number of pages)

Pages 6754-6762 (9 pages)

Journal (Volume, Issue Number)

Medical Physics (Volume 38, Issue 12)

Publication milestones

  • Published - 12/2011

Publication status

Published - 12/2011

ISSN

0094-2405

Publication IDs

  • Scopus: 82755163790
  • PubMed: 22149857

Publication metrics

Metrics

SciVal
FWCI
2.31
SciVal
Author count
7
SciVal
citations
58
SciVal
Paper percentile
93
SciVal
Top percentile
10
Fractional count
1
Fractional count
0.14
Fractional count
6
Fractional count
0.86
Fractional count
1
Fractional count
1
Scopus
citations

PlumX, opens in new tab

Captures
65
Citation count
76

Funding Details

This work was supported by NIH R01CA100835, S103034539, NCI S10RR027610, NIAID K02AI093866, and NASA NSCOR NNX11AC60G.
FundersFunding numbers
NIH
S103034539, R01CA100835
NCI
S10RR027610
NIAID
K02AI093866
NASA
NNX11AC60G