Commissioning and benchmarking a 3D dosimetry system for clinical use

  • Andrew Thomas
    ,
  • ,
  • John Adamovics
    ,
  • Mark Oldham(corresponding author)
*Corresponding author for this work
  • Duke University
    ,
  • Rider University
Scholary Output:
Contribution to journal
Article
Peer-review

Open access

Abstract

Purpose: A 3D dosimetry system is described which consists of two parts: a radiochromic plastic dosimeter PRESAGE® (which responds to absorbed dose with a linear change in optical-density) and the Duke large-field-of-view optical-CT scanner (DLOS). The DLOS/PRESAGE system has recently been commissioned and benchmarked for clinical use and, in particular, for verification and commissioning of complex radiation treatments. Methods: DLOS commissioning involved determining the dynamic range, spatial resolution, noise, temporal, and other characteristics of the light source and imaging components. Benchmarking tests were performed on the combined DLOS/PRESAGE system to establish baseline dosimetric performance. The tests consisted of delivering simple radiation treatments to PRESAGE dosimeters, and comparing the measured 3D relative dose distributions with the known gold standard. The gold standard distribution was obtained from machine beam-data or the treatment planning system (TPS). All studies used standardized procedures to ensure consistency. Results: For commissioning, isotropic spatial resolution was submillimeter (MTF >0.5 for frequencies of 1.5 lp/mm) and the dynamic range was ∼60 dB. Flood field uniformity was within 10% and stable after 45 min of warm-up. Stray-light is small, due to telecentricity, but even the residual can be removed through deconvolution by a point-spread-function. For benchmarking, the mean 3D passing NDD (normalized dose distribution) rate (3%, 3mm, 5% dose threshold) over the benchmark data sets was 97.3% ± 0.6% (range 96%-98%), which is on par with other planar dosimeters used in external beam radiation therapy indicating excellent agreement. Noise was low at <2% of maximum dose (4-12 Gy) for 2 mm reconstructions. The telecentric design was critical to enabling fast imaging with minimal stray-light artifacts. Conclusions: This work presents the first comprehensive benchmarking of a 3D dosimetry system for clinical use. The DLOS/PRESAGE benchmark tests show consistently good agreement to simple known distributions. The system produces accurate isotropic 2 mm dose data over clinical volumes (e.g., 16 cm diameter phantoms, 12 cm height), in under 15 min. It represents a uniquely useful and versatile new tool for commissioning and verification of complex therapy treatments.

Publication Information

Output type

Scholary Output:
Contribution to journal
Article
Peer-review

Original language

English (US)

Pages from-to (Number of pages)

Pages 4846-4857 (12 pages)

Journal (Volume, Issue Number)

Medical Physics (Volume 38, Issue 8)

Publication milestones

  • Published - 08/2011

Publication status

Published - 08/2011

ISSN

0094-2405

Publication IDs

  • Scopus: 79961085442
  • PubMed: 21928656

Publication metrics

Metrics

SciVal
FWCI
4.63
SciVal
Author count
4
SciVal
citations
77
SciVal
Paper percentile
95
SciVal
Top percentile
5
Scopus
citations
Fractional count
1
Fractional count
0.25
Fractional count
3
Fractional count
0.75
Fractional count
1
Fractional count
1

PlumX, opens in new tab

Citation count
87
Captures
33

Funding Details

FunderFunding number
H2020
676629