Continuous spectrophotometric measurements of arteriovenous oxygen difference
- A. P. Shepherd,
- ,
- A. F. Wilson
- University of Texas Health Science Center at San Antonio
Abstract
A relatively inexpensive dual beam optical absorbance monitor was tested and adapted to measure arteriovenous oxygen difference ((a-v)ΔO2). Since the molar extinction coefficients of oxy and deoxyhemoglobin are markedly different at 660 nm, the difference in optical density between arterial and venous blood, as measured by the absorbance monitor, was shown to be linearly related to (a-v)ΔO2. When appropriate cuvettes (optical path length = 0.5 mm) were used, the instrument's 90% response time (4.4 sec) was sufficiently rapid for most physical applications such as Fick determinations of the rate of an organ's oxygen consumption. The theoretical basis of this spectrophotometric (a-v)ΔO2 measurement and sources of error are discussed in detail.
Publication Information
Output type
Original language
English (US)Pages from-to (Number of pages)
Pages 152-155 (4 pages)Journal (Volume, Issue Number)
Journal of Applied Physiology (Volume 39, Issue 1)Publication milestones
- Published - 1975
Publication status
ISSN
8750-7587Publication IDs
- Scopus: 0016820842
- PubMed: 1150582