Convolution-subtraction technique to compensate for downscatter effects in simultaneous Tl-201/Tc-99m cardiac SPECT
- Lin Qian(corresponding author),
- Zong Jian Cao
- University of Maryland, Baltimore,
Scholary Output:
Contribution to conference
Paper
Peer-reviewRelated Event
Title
Proceedings of the 1997 IEEE Nuclear Science Symposium
Event type
OtherDate
11/09/1997 - 11/15/1997Location
Albuquerque, NM, USA
Abstract
A convolution-subtraction method to remove the downscatter counts in the 70 keV projection data is proposed. First, the scatter response function (SRF) is derived by fitting the Monte Carlo simulated downscatter data acquired in the 70 keV window for Tc-99m point source located in a uniform elliptical cylinder. Then, the derived SRF is convolved with the reconstructed 140 keV image to obtain the spatial distribution of downscatter counts in the 70 keV window. Finally, the downscatter distribution is subtracted from the 70 keV data to obtain the compensated data.