Correction of Bowtie-Filter Normalization and Crescent Artifacts for a Clinical CBCT System

  • Hong Zhang
    ,
  • Vic Kong
    ,
  • Ke Huang
    ,
  • Jian Yue Jin(corresponding author)
*Corresponding author for this work
  • Augusta University
Scholary Output:
Contribution to journal
Article
Peer-review

Sustainable Development Goals

  • SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well

Abstract

Purpose: To present our experiences in understanding and minimizing bowtie-filter crescent artifacts and bowtie-filter normalization artifacts in a clinical cone beam computed tomography system. Methods: Bowtie-filter position and profile variations during gantry rotation were studied. Two previously proposed strategies (A and B) were applied to the clinical cone beam computed tomography system to correct bowtie-filter crescent artifacts. Physical calibration and analytical approaches were used to minimize the norm phantom misalignment and to correct for bowtie-filter normalization artifacts. A combined procedure to reduce bowtie-filter crescent artifacts and bowtie-filter normalization artifacts was proposed and tested on a norm phantom, CatPhan, and a patient and evaluated using standard deviation of Hounsfield unit along a sampling line. Results: The bowtie-filter exhibited not only a translational shift but also an amplitude variation in its projection profile during gantry rotation. Strategy B was better than strategy A slightly in minimizing bowtie-filter crescent artifacts, possibly because it corrected the amplitude variation, suggesting that the amplitude variation plays a role in bowtie-filter crescent artifacts. The physical calibration largely reduced the misalignment-induced bowtie-filter normalization artifacts, and the analytical approach further reduced bowtie-filter normalization artifacts. The combined procedure minimized both bowtie-filter crescent artifacts and bowtie-filter normalization artifacts, with Hounsfield unit standard deviation being 63.2, 45.0, 35.0, and 18.8 Hounsfield unit for the best correction approaches of none, bowtie-filter crescent artifacts, bowtie-filter normalization artifacts, and bowtie-filter normalization artifacts + bowtie-filter crescent artifacts, respectively. The combined procedure also demonstrated reduction of bowtie-filter crescent artifacts and bowtie-filter normalization artifacts in a CatPhan and a patient. Conclusion: We have developed a step-by-step procedure that can be directly used in clinical cone beam computed tomography systems to minimize both bowtie-filter crescent artifacts and bowtie-filter normalization artifacts.

Publication Information

Output type

Scholary Output:
Contribution to journal
Article
Peer-review

Original language

English (US)

Pages from-to (Number of pages)

Pages 81-91 (11 pages)

Journal (Volume, Issue Number)

Technology in Cancer Research and Treatment (Volume 16, Issue 1)

Publication milestones

  • Published - 02/2017

Publication status

Published - 02/2017

ISSN

1533-0346

Publication IDs

  • Scopus: 85011591474
  • PubMed: 26834116

Publication metrics

Metrics

Scopus
citations
SciVal
Author count
4
SciVal
Paper percentile
25
Fractional count
2
Fractional count
0.50
Fractional count
2
Fractional count
0.50
Fractional count
2
Fractional count
1

PlumX, opens in new tab

Citation count
7
Captures
18

Funding Details

The author(s) disclosed receipt of the following financial support for the research, authorship, and/or publication of this article: This study is supported by Award Number 1R01CA166948 from the NIH and National Cancer Institute.