Discordance between viral loads determined by Roche COBAS AMPLICOR human immunodeficiency virus type 1 monitor (version 1.5) standard and ultrasensitive assays caused by freezing patient plasma in centrifuged Becton-Dickinson vacutainer brand plasma preparation tubes

  • Hossein Salimnia(corresponding author)
    ,
  • Ellen C. Moore
    ,
  • Lawrence R. Crane
    ,
  • Rodger David MacArthur
    ,
  • Marilynn R. Fairfax
*Corresponding author for this work
  • Wayne State University
    ,
  • DMC University Laboratories
Scholary Output:
Contribution to journal
Article
Peer-review

Sustainable Development Goals

  • SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well

Abstract

The Roche COBAS AMPLICOR human immunodeficiency virus type 1 (HIV-1) Monitor (version 1.5) standard and ultrasensitive viral load assays often gave discordant results, with viral loads from the standard assay exceeding those from the ultrasensitive assay by more than 0.5 log10 for approximately 20% of specimens received. We began studies to determine the extent, magnitude, and reproducibility of the discordance between the assays and to discover and eliminate the cause of this discordance. Until then, we revised our standard operating procedure to include both standard and ultrasensitive testing on all specimens submitted for viral load determinations. Discordant results usually recurred on retesting. They were most prevalent for specimens with ultrasensitive viral loads of <1,000 and rare for specimens with viral loads of >10,000. Often, standard assay results exceeded those of the ultrasensitive assay by 50- to 100-fold. At higher viral loads, the difference between the standard and ultrasensitive assays persisted, but the percent difference was smaller and rarely caused discordance. The proportion of discordant results was significantly higher in specimens from pediatric patients than in specimens from adults. The ultrasensitive viral load determinations generally agreed with the results of the B-DNA (Bayer) viral load assays. If the plasma was transferred from the centrifuged plasma preparation tubes before freezing, standard and ultrasensitive results were concordant with each other and with values determined on plasma from lavender-topped EDTA tubes.

Publication Information

Output type

Scholary Output:
Contribution to journal
Article
Peer-review

Original language

English (US)

Pages from-to (Number of pages)

Pages 4635-4639 (5 pages)

Journal (Volume, Issue Number)

Journal of Clinical Microbiology (Volume 43, Issue 9)

Publication milestones

  • Published - 09/01/2005

Publication status

Published - 09/01/2005

ISSN

0095-1137

Publication IDs

  • Scopus: 24744446821
  • PubMed: 16145119
  • Scopus: 24744446821

Publication metrics

Metrics

Scopus
citations
SciVal
citations
21
SciVal
FWCI
0.44
SciVal
Author count
5
SciVal
Paper percentile
73
Fractional count
1
Fractional count
0.20
Fractional count
4
Fractional count
0.80
Fractional count
1
Fractional count
1

PlumX, opens in new tab

Citation count
32
Captures
16