Feasibility of a reusable radiochromic dosimeter

*Corresponding author for this work
Scholary Output:
Contribution to journal
Article
Research
Peer-review

Open access

Abstract

The current practice for patient-specific quality assurance (QA) uses ion chambers or diode arrays primarily because of their ease of use and reliability. A standard routine compares the dose distribution measured in a phantom with the dose distribution calculated by the treatment planning system for the same experimental conditions. For the particular problems encountered in the treatment planning of complex radiotherapy techniques, such as small fields/segments and dynamic delivery systems, additional tests are required to verify the accuracy of dose calculations. The dose distribution verification should be throughout the total 3D dose distribution for a high dose gradient in a small, irradiated volume, instead of the standard practice of one to several planes with 2D radiochromic (GAFChromic) film. To address this issue, we have developed a 3D radiochromic dosimeter that improves the rigor of current QA techniques by providing high-resolution, complete 3D verification for a wide range of clinical applications. The dosimeter is composed of polyurethane, a radical initiator, and a leuco dye, which is radiolytically oxidized to a dye absorbing at 633 nm. Since this chemical dosimeter is single-use, it represents a significant expense. The purpose of this research is to develop a cost-effective reusable dosimeter formulation. Based on prior reusability studies, three promising dosimeter formulations were studied using small volume optical cuvettes and irradiated to known clinically relevant doses of 0.5–10 Gy. After irradiation, the change in optical density was measured in a spectrophotometer. All three formulations retained linearity of optical density response to radiation upon re-irradiations. However, only one formulation retained dose sensitivity upon at least five re-irradiations, making it ideal for further evaluation as a 3D dosimeter.

Publication Information

Output type

Scholary Output:
Contribution to journal
Article
Research
Peer-review

Original language

English (US)

Article number

9906

Journal (Volume, Issue Number)

Applied Sciences (Switzerland) (Volume 11, Issue 21)

Publication milestones

  • Published - 11/01/2021

Publication status

Published - 11/01/2021

Publication IDs

  • Scopus: 85117597147

Publication metrics

Metrics

Scopus
citations
Fractional count
2
Fractional count
0.29
Fractional count
5
Fractional count
0.71
Fractional count
2
Fractional count
1

PlumX, opens in new tab

Captures
8
Citation count
4