Skip to search boxSkip to navigationSkip to main content

Heat generation during implant drilling: The significance of motor speed

  • Mohamed Sharawy(corresponding author)
    ,
  • Carl E. Misch
    ,
  • Norman Weller
    ,
  • Sherif Tehemar
*Corresponding author for this work
Scholary Output:
Contribution to journal
Article
Research
Peer-review

Abstract

<p>Purpose: The purpose of this study was to measure the heat generated from 3 drilling speeds (1,225, 1,667, and 2,500 rpm) using the armamentarium of 4 implant systems. Materials and Methods: The mean rise in temperature, the time of drilling, and the time needed for pig jaw bone to return to the baseline temperature were monitored using 4 thermocouple technology. Results: The mean rise in temperature, the time of drilling, and the time needed for the specimens to return to the baseline temperature were lower at 2,500 rpm than at 1,667 or 1,225 rpm (P ≤ .05), regardless of the system used. The rpm also directly correlated to the amount of time the bone remained at an elevated temperature. Conclusion: From a heat generation standpoint, we conclude that preparing an implant site at 2500 rpm could decrease the risk of osseous damage, which may affect the initial healing of dental implants. This may decrease the devital zone adjacent to an implant after surgery and be most advantageous in immediate load application to dental implants.</p>

Publication Information

Output type

Scholary Output:
Contribution to journal
Article
Research
Peer-review

Original language

English (US)

Article number

00094

Pages from-to (Number of pages)

Pages 1160-1169 (10 pages)

Journal (Volume, Issue Number)

Journal of Oral and Maxillofacial Surgery (Volume 60, Issue 10)

Publication milestones

  • Published - 2002

Publication status

Published - 2002

ISSN

0278-2391

Publication IDs

  • Scopus: 0036788376
  • PubMed: 12378492

Publication metrics

Metrics

SciVal
citations
159
Fractional count
1
Fractional count
0.25
Fractional count
3
Fractional count
0.75
Fractional count
1
Fractional count
1
SciVal
FWCI
1.55
SciVal
Author count
4
SciVal
Paper percentile
96
SciVal
Top percentile
5
Scopus
citations

PlumX, opens in new tab

Citation count
210
Mentions
1
Captures
205

Funding Details

The study was funded in part by the Misch Implant Institute, Birmingham, MI.
FunderFunding numbers
Misch Implant Institute
-