Heparin coating of vascular prostheses reduces thromboemboli
- Edmond F. Ritter,
- Yong Bae Kim,
- Helmut P. Reischl,
- Donald Serafin,
- Adam M. Rudner,
- Bruce Klitzman(corresponding author)
- Unknown,
- Duke University
Abstract
Background. Synthetic conduits made from currently available materials are suboptimal for use in small-diameter vascular reconstruction because of their high surface thrombogenicity, which leads to failure. Methods. In this study control, heparin-irrigated, or heparin-bonded expanded polytetrafluoroethylene (ePTFE) grafts (4 mm long by 1 mm inner diameter) were implanted to reconstruct the iliac artery in male rats. The cremaster muscle was isolated as an island flap based on branches of the iliac artery downstream from the graft. Emboli were quantitated by using intravital fluorescent microscopy of the cremaster muscle's microcirculation. Results. The mean number of emboli observed per animal during a 20-minute period was 91 for the control group, 84 for the heparin-irrigated group, and 22 for the tridodecylmethylammonium chloride (TDMAC)-heparin group. The mean area of each embolus was 1057μm2 for control, 940 μm2 for heparin-irrigated, and 808 μm2 for TDMAC-heparin-coated grafts (p < 0.05 for TDMAC-heparin versus control or heparin-irrigated). Conclusions. A TDMAC-heparin coating of ePTFE microvascular prostheses significantly reduces downstream microemboli.
Publication Information
Output type
Original language
English (US)Pages from-to (Number of pages)
Pages 888-892 (5 pages)Journal (Volume, Issue Number)
Surgery (Volume 122, Issue 5)Publication milestones
- Published - 11/1997
Publication status
ISSN
0039-6060Publication IDs
- Scopus: 12644313246
- PubMed: 9369888