In vitro and in vivo measurement of remaining dentin thickness

  • John F. Hatton
    ,
  • David H. Pashley
    ,
  • James Shunk
    ,
  • Gregory P. Stewart
  • Unknown
Scholary Output:
Contribution to journal
Article
Peer-review

Abstract

Experiments were performed in vitro on extracted human teeth and in vivo on dog teeth to determine the accuracy of a pulse-echo mode with an ultrasonic micrometer in measuring remaining dentin thickness. The accuracy of the ultrasonic micrometer measured remaining dentin thickness was compared with direct caliper measurements in measuring human dentin discs (correlation coefficient, r = 0.99). After modification of the delay line tip (focusing tip), the device was found to be highly accurate on both prepared extracted human teeth (r = 0.87) and in vivo on dog teeth (r = 0.92). The use of ultrasonic technology to measure remaining dentin thickness looks promising. Further research needs to be performed to make the device more clinically useful.

Publication Information

Output type

Scholary Output:
Contribution to journal
Article
Peer-review

Original language

English (US)

Pages from-to (Number of pages)

Pages 580-584 (5 pages)

Journal (Volume, Issue Number)

Journal of endodontics (Volume 20, Issue 12)

Publication milestones

  • Published - 12/1994

Publication status

Published - 12/1994

ISSN

0099-2399

Publication IDs

  • Scopus: 0028723881
  • PubMed: 7759982

Publication metrics

Metrics

Scopus
citations
Fractional count
1
Fractional count
0.25
Fractional count
3
Fractional count
0.75
Fractional count
1
Fractional count
1

PlumX, opens in new tab

Captures
23
Citation count
21