Influence of Class V preparation on in vivo temperature rise in anesthetized human pulp during exposure to a Polywave® LED light curing unit

  • Driellen Christine Zarpellon
    ,
  • Patrício Runnacles
    ,
  • Cristiane Maucoski
    ,
  • Dayane Jaqueline Gross
    ,
  • Ulisses Coelho
    ,
*Corresponding author for this work
Scholary Output:
Contribution to journal
Article
Peer-review

Abstract

Objective: This in vivo study evaluated pulp temperature (PT) rise in human premolars having deep Class V preparations during exposure to a light curing unit (LCU) using selected exposure modes (EMs). Methods: After local Ethics Committee approval, intact first premolars (n = 8) requiring extraction for orthodontic reasons, from 8 volunteers, received infiltrative and intraligamental anesthesia and were isolated using rubber dam. A minute pulp exposure was attained and sterile probe from a wireless, NIST-traceable, temperature acquisition system was inserted into the coronal pulp chamber to continuously monitor PT (°C). A deep buccal Class V preparation was prepared using a high speed diamond bur under air-water spray cooling. The surface was exposed to a Polywave® LED LCU (Bluephase 20i, Ivoclar Vivadent) using selected EMs, allowing 7-min span between each exposure: 10-s in low (10-s/L), 10-s (10-s/H), 30-s (30-s/H), or 60-s (60-s/H) in high mode; and 5-s-Turbo (5-s/T). Peak PT values and PT increases over physiologic baseline levels (ΔT) were subjected to 1-way, repeated measures ANOVAs, and Bonferroni's post-hoc tests (α = 0.05). Linear regression analysis was performed to establish the relationship between applied radiant exposure and ΔT. Results: All EMs produced higher peak PT than the baseline temperature (p < 0.001). Only 60-s/H mode generated an average ΔT of 5.5 °C (p < 0.001). A significant, positive relationship was noted between applied radiant exposure and ΔT (r2 = 0.8962; p < 0.001). Significance: In vivo exposure of deep Class V preparation to Polywave® LED LCU increases PT to values considered safe for the pulp, for most EMs. Only the longest evaluated EM caused higher PT increase than the critical ΔT, thought to be associated with pulpal necrosis.

Publication Information

Output type

Scholary Output:
Contribution to journal
Article
Peer-review

Original language

English (US)

Pages from-to (Number of pages)

Pages 901-909 (9 pages)

Journal (Volume, Issue Number)

Dental Materials (Volume 34, Issue 6)

Publication milestones

  • Accepted/In press - 01/01/2018
  • Published - 06/2018

Publication status

Published - 06/2018

ISSN

0109-5641

Publication IDs

  • Scopus: 85044535073
  • ORCID: /0000-0003-1947-3337/work/54796699
  • PubMed: 29606368

Publication metrics

Metrics

SciVal
FWCI
0.48
SciVal
Author count
7
SciVal
citations
4
SciVal
Paper percentile
62
Scopus
citations
Fractional count
1
Fractional count
0.14
Fractional count
6
Fractional count
0.86
Fractional count
1
Fractional count
1

PlumX, opens in new tab

Captures
50
Citation count
27

Funding Details

The authors are indebted to the State University of Ponta Grossa and Augusta University for support required to perform this study. The authors would also like to thank Ivoclar Vivadent Inc. for donation of the light curing unit. This study was financially supported by grants #232/2014 and #488/2014 from Araucaria Foundation, Parana, Brazil .
FunderFunding numbers
Araucaria Foundation
-