Investigating the reproducibility of a complex multifocal radiosurgery treatment
- M. Niebanck,
- T. Juang,
- ,
- J. Adamovics,
- Z. Wang,
- M. Oldham
- Duke University,
- Rider University
Open access
Related Event
Title
Event type
OtherDate
12/04/2012 - 12/08/2012Location
Abstract
Stereotactic radiosurgery has become a widely used technique to treat solid tumors and secondary metastases of the brain. Multiple targets can be simultaneously treated with a single isocenter in order to reduce the set-up time to improve patient comfort and workflow. In this study, a 5-arc multifocal RapidArc treatment was delivered to multiple PRESAGE® dosimeters in order to explore the repeatability of the treatment. The three delivery measurements agreed well with each other, with less than 3% standard deviation of dose in the target. The deliveries also agreed well with the treatment plan, with gamma passing rates greater than 90% (5% dose-difference, and 2 mm distance-to-agreement criteria). The optical-CT PRESAGE® system provided a reproducible measurement for treatment verification, provided measurements were made immediately following treatment.
Publication Information
Output type
Original language
English (US)Article number
012072Journal (Volume, Issue Number)
Journal of Physics: Conference Series (Volume 444, Issue 1)Publication milestones
- Published - 2013
Publication status
ISSN
1742-6588Publication IDs
- Scopus: 84883374951