Light flux density threshold at which protein denaturation is induced by synchrotron radiation circular dichroism beamlines
- A. J. Miles,
- Robert W. Janes,
- A. Brown,
- D. T. Clarke,
- ,
- Y. Tao
- Birkbeck University of London,
- Queen Mary University of London,
- Daresbury Laboratory,
- East Carolina University,
- Brookhaven National Laboratory,
- CAS - Institute of High Energy Physics
Scholary Output:
Contribution to journal
Article
Peer-reviewOpen access
Abstract
New high-flux synchrotron radiation circular dichroism (SRCD) beamlines are providing important information for structural biology, but can potentially cause denaturation of the protein samples under investigation. This effect has been studied at the new CD1 dedicated SRCD beamline at ISA in Denmark, where radiation-induced thermal damage effects were observed, depending not only on the radiation flux but also on the focal spot size of the light. Comparisons with similar studies at other SRCD facilities worldwide has lead to the estimation of a flux density threshold under which SRCD beamlines should be operated when samples are to be exposed to low-wavelength vacuum ultraviolet radiation for extended periods of time.
Publication Information
Output type
Scholary Output:
Contribution to journal
Article
Peer-reviewOriginal language
English (US)Pages from-to (Number of pages)
Pages 420-422 (3 pages)Journal (Volume, Issue Number)
Journal of Synchrotron Radiation (Volume 15, Issue 4)Publication milestones
- Published - 2008
Publication status
Published - 2008
ISSN
0909-0495Publication IDs
- Scopus: 46149090409
- PubMed: 18552437
Publication metrics
Metrics
SciVal
FWCI
2.90
SciVal
Author count
8
SciVal
citations
74
SciVal
Paper percentile
93
SciVal
Top percentile
10
Fractional count
1
Fractional count
0.13
Fractional count
7
Fractional count
0.88
Fractional count
1
Fractional count
1
PlumX, opens in new tab
Mentions
1
Captures
27
Citation count
93