Measurement of renal tissue oxygen tension: Systematic differences between fluorescence optode and microelectrode recordings in anaesthetized rabbits
- Chai Ling Leong,
- ,
- Gabriela A. Eppel,
- Warwick P. Anderson,
- Roger G. Evans
- Monash University,
- Medical College of Wisconsin
Scholary Output:
Contribution to journal
Article
Research
Peer-reviewAbstract
<p>Background/Aims: The validity of fluorescence optodes for measurement of renal cortical tissue oxygen tension was tested by comparison with Clark electrodes. Methods: We varied renal blood flow and inspired O<sub>2</sub> content in anaesthetized rabbits while simultaneously measuring cortical tissue oxygen tension. Results: Cortical oxygen tension varied with inspired O <sub>2</sub> content. Fluorescence optode measurements were more tightly distributed than those from a Clark electrode. Cumulative frequency distributions for fluorescence optodes were shifted to the left of those for Clark electrodes. The slope of the relationship between oxygen tension in arterial blood and cortical tissue was less for the fluorescence optode than the Clark electrode. Cortical tissue oxygen tension measurements by these two methods were correlated (r<sup>2</sup> = 0.32; p < 0.001), with no fixed bias but considerable proportional bias. Thus, the slope of the relationship between the two measurements was less than unity (0.57 [0.50-0.69]). Conclusion: Cortical oxygen tension values from fluorescence optodes are less variable but proportionally less than those from Clark electrodes. Theoretical considerations suggest that true interstitial oxygen tension lies somewhere between values provided by the two techniques. Nevertheless, the lesser variability of the fluorescence optode technique may aid detection of physiologically significant changes in intrarenal oxygenation.</p>
Publication Information
Output type
Scholary Output:
Contribution to journal
Article
Research
Peer-reviewOriginal language
English (US)Pages from-to (Number of pages)
Pages p11-p17Journal (Volume, Issue Number)
Nephron - Physiology (Volume 108, Issue 2)Publication milestones
- Published - 03/2008
Publication status
Published - 03/2008
ISSN
1660-2137Publication IDs
- Scopus: 40449095968
- PubMed: 18223307
Publication metrics
Metrics
SciVal
citations
20
Fractional count
1
Fractional count
0.20
Fractional count
4
Fractional count
0.80
Fractional count
1
Fractional count
1
SciVal
FWCI
0.88
SciVal
Author count
5
SciVal
Paper percentile
74
PlumX, opens in new tab
Captures
12
Citation count
27