Minimizing interference and false alarms from electrocautery in the Nellcor N-100 pulse oximeter

  • Frank E Block(corresponding author)
    ,
  • George J. Detko
*Corresponding author for this work
  • Indian River Memorial Hospital
Scholary Output:
Contribution to journal
Article
Research
Peer-review

Abstract

Because the Nellcor N-100 pulse oximeter is generally quite sensitive to electrocautery interference, erroneous oxygen saturation values may be displayed and an audible false alarm may sound. We have discovered, through trial and error, several techniques that may minimize both the interference and the false alarms. Keeping the pulse oximeter sensor and the oximeter unit as far as possible from the surgical site and the electrocautery grounding plate will usually decrease the interference. If this step is inadequate, the incidence of false alarms can be decreased by setting the high pulse rate alarm to approximately 160, operating the unit in mode 2 (the rapid-response mode), and setting the "alarm off" function to 120 seconds. These steps will usually markedly decrease the incidence of false alarms.

Publication Information

Output type

Scholary Output:
Contribution to journal
Article
Research
Peer-review

Original language

English (US)

Pages from-to (Number of pages)

Pages 203-205 (3 pages)

Journal (Volume, Issue Number)

Journal of Clinical Monitoring (Volume 2, Issue 3)

Publication milestones

  • Published - 07/01/1986

Publication status

Published - 07/01/1986

ISSN

0748-1977

Publication IDs

  • Scopus: 0022511724
  • PubMed: 3746371

Publication metrics

Metrics

Fractional count
1
Fractional count
0.50
Fractional count
1
Fractional count
0.50
Fractional count
1
Fractional count
1
Scopus
citations

PlumX, opens in new tab

Citation count
13
Captures
2