Monte Carlo analysis of beam blocking grid design parameters: Scatter estimation and the importance of electron backscatter: Scatter

  • Gregory J. Bootsma(corresponding author)
    ,
  • Lei Ren
    ,
  • Hong Zhang
    ,
  • JianYue Jin
    ,
  • David A. Jaffray
*Corresponding author for this work
  • University Health Network
    ,
  • Duke University
    ,
  • Indiana University-Purdue University Indianapolis
    ,
  • ,
  • University of Toronto
Scholary Output:
Contribution to journal
Article
Peer-review

Open access

Abstract

Purpose: Beam blocking grids provide a simple and direct measurement of the scattered photon signal which degrades image quality in x-ray imaging systems, such as cone-beam CT (CBCT). This study evaluates the scatter estimation accuracy of the beam blocking method to optimize the design parameters of the grid system (e.g., grid thickness, source-To-grid distance (SGD), septa width, air interspace, and grid ratio) using Monte Carlo (MC) simulations. Method: A MC model of a CBCT imaging system with a beam blocking grid in place is made using code based on EGSnrc, with the x-ray tube portion of the simulation including electron backscatter between the anode and cathode. The inclusion of the electron backscatter allowed a more complete model of the contamination signal to be estimated. The contamination signal consists of the off-focal radiation (OFR) and source component scatter (photon scatter in source components such as tube housing, filters, and collimators). The MC model was validated against measurements collected on a bench top imaging system with a grid in place. The MC model was used to simulate 11 different grid design configurations in addition to a case with no grid. For each design a simulated projection with and without a phantom in place was computed. The simulated projections were then used to estimate the scatter and contamination portion of the signal using the signal behind the grid septa. The estimated signals from the grid data were compared to the actual signals labeled during the MC simulation. Results: Simulated results showed good agreeance with measured results with the importance of including electron backscatter resulting in off-focal radiation in the simulation being highlighted. When the source was free of contamination photons all grids performed with an error less than 8% when estimating just the scatter from the object. When the contamination photons were included in the simulation, the error in estimating both the scatter and contamination signal rose by a factor of 4 on average. In the case when both signals are present, increasing the grid thickness, changing the SGD, and reducing septa width and air interspace sizes all showed the ability to improve the grid-based estimates of the object scatter and contamination portion signal. Conclusions: The inclusion of the contamination signal in MC simulations of x-ray imaging systems is important in the design, validation, and evaluation of measurement-based scatter methods. Beam blocking grids show potential not only in object scatter estimation but in the estimation of the contamination signal, but appropriate interpolation functions must be used to account for higher frequencies found in contamination signal.

Publication Information

Output type

Scholary Output:
Contribution to journal
Article
Peer-review

Original language

English (US)

Pages from-to (Number of pages)

Pages 1059-1070 (12 pages)

Journal (Volume, Issue Number)

Medical Physics (Volume 45, Issue 3)

Publication milestones

  • Accepted/In press - 01/01/2018
  • Published - 03/2018

Publication status

Published - 03/2018

ISSN

0094-2405

Publication IDs

  • Scopus: 85041722753

Publication metrics

Metrics

SciVal
FWCI
0.26
SciVal
Author count
5
SciVal
citations
2
SciVal
Paper percentile
49
Scopus
citations
Fractional count
1
Fractional count
0.20
Fractional count
4
Fractional count
0.80
Fractional count
1
Fractional count
1

PlumX, opens in new tab

Citation count
5
Captures
18