Skip to search boxSkip to navigationSkip to main content

Rotablation Wire Fracture During a Challenging Percutaneous Coronary Intervention

*Corresponding author for this work
  • Augusta University
    ,
  • 424 General Military Training Hospital, Greece
    ,
Scholary Output:
Contribution to journal
Article
Research
Peer-review

Open access

Abstract

<p>Background: Rotablation is considered a standard plaque-modification procedure in the treatment of calcified coronary lesions. The rotablation coronary wire is specifically designed to endure the high pressures and temperatures developed during rotablation. Case Summary: This is a case report of a severely calcified proximal right coronary artery coronary lesion rotablation. Several runs were successfully performed at the lesion; however, during the final run, the rotablation wire was severed and entrapped at the lesion site. It was snared and partially retrieved out of the vessel. The intravascular ultrasound–guided percutaneous coronary intervention was completed successfully. Discussion: Only 2 other reported cases of rotablation wire fracture exist. Despite the specific design of the rotablation wire, several factors (eg, heavy calcification, difficult anatomy, and “wire fatigue”) seem to have caused this unusual complication. Take-Home Messages: Even specifically designed, “unbreakable” wires may sometimes break. Usually complex cases do not go as planned, so the operator needs to “improvise, adjust, and overcome.”</p>

Publication Information

Output type

Scholary Output:
Contribution to journal
Article
Research
Peer-review

Original language

English (US)

Article number

104745

Journal (Volume, Issue Number)

JACC: Case Reports (Volume 30, Issue 24)

Publication milestones

  • Published - 08/20/2025

Publication status

Published - 08/20/2025

Publication IDs

  • Scopus: 105012768601

Publication metrics

Metrics

Fractional count
2
Fractional count
0.40
Fractional count
3
Fractional count
0.60
Fractional count
2
Fractional count
1