Securing extraocular muscles in strabismus surgery: effect of cautery on suture strength
- Carolina Adams,
- Steven E. Brooks(corresponding author)
- Columbia University
Scholary Output:
Contribution to journal
Article
Peer-reviewAbstract
An experimental study was performed to evaluate quantitatively the effect of cautery on the tensile strength of sutures commonly used in strabismus surgery. This in vitro study was conducted in a controlled fashion using 6-0 polyglactin 910 suture, two different forms of cautery, and a precision digital force gauge. The results suggest that thermal electrocautery with a wire tip can substantially weaken or break 6-0 polyglactin threads only if direct contact is made, but bipolar cautery at typical ophthalmic surgical settings does not.