SU‐D‐BRCD‐06: Measurement of Elekta Electron Energy Spectra Using a Small Magnetic Spectrometer

  • K. Hogstrom
    ,
  • D. Mclaughlin
    ,
  • J. Gibbons
    ,
  • P. Shikhaliev
    ,
  • T. Clarke
    ,
  • A. Henderson
  • Louisiana State University
    ,
  • Mary Bird Perkins Cancer Center
    ,
  • Rice University
Scholary Output:
Contribution to journal
Article
Peer-review

Abstract

Purpose: To demonstrate how a small magnetic spectrometer can measure the energy spectra of seven electron beams on an Elekta Infinity tuned to match beams on a previously commissioned machine. Methods: Energyspectra were determined from measurements of intensity profiles on 6″‐long computed radiographic (CR) strips after deflecting a narrow incident beam using a small (28 lbs.), permanent magnetic spectrometer. CR plateexposures (<1cGy) required special beam reduction techniques and bremsstrahlung shielding. Curves of CR intensity (corrected for non‐ linearity and background) versus position were transformed into energy spectra using the transformation from position (x) on the CR plate to energy (E) based on the Lorentz force law. The effective magnetic field and its effective edge, parameters in the transformation, were obtained by fitting a plot of most probable incident energy (determined from practical range) to the peak position. Results: The calibration curve (E vs. x) fit gave 0.423 Tesla for the effective magnetic field. Most resulting energy spectra were characterized by a single, asymmetric peak with peak position and FWHM increasing monotonically with beam energy. Only the 9‐MeV spectrum was atypical, possibly indicating suboptimal beam tuning. These results compared well with energy spectra independently determined by adjusting each spectrum until the EGSnrc Monte Carlo calculated percent depth‐dose curve agreed well with the corresponding measured curve. Conclusions: Results indicate that this spectrometer and methodology could be useful for measuring energy spectra of clinical electron beams at isocenter. Future work will (1) remove the small effect of the detector response function (due to pinhole size and incident angular spread) from the energy spectra, (2) extract the energy spectra exiting the accelerator from current results, (3) use the spectrometer to compare energy spectra of matched beams among our clinical sites, and (4) modify the spectrometer to utilize radiochromic film.

Publication Information

Output type

Scholary Output:
Contribution to journal
Article
Peer-review

Original language

English (US)

Pages from-to (Number of pages)

Page 3614 (1 page)

Journal (Volume, Issue Number)

Medical Physics (Volume 39, Issue 6)

Publication milestones

  • Published - 06/2012

Publication status

Published - 06/2012

ISSN

0094-2405

Publication IDs

  • Scopus: 85024806255

Publication metrics

Metrics

SciVal
Author count
9
SciVal
Paper percentile
24
Fractional count
1
Fractional count
0.11
Fractional count
8
Fractional count
0.89
Fractional count
1
Fractional count
1