Time-resolved fluorescence polarization measurements for entire emission spectra with a resistive-anode, single-photon-counting detector: The Fluorescence omnilyzer

*Corresponding author for this work
  • University of Maryland, Baltimore County
    ,
  • Brookhaven National Laboratory
Scholary Output:
Contribution to journal
Article
Peer-review

Abstract

We report a fluorescence analyzer that records simultaneously the temporal profiles for both orthogonal linear polarizations for all wavelengths in a fluorescence emission spectrum. The Analyzer combines a resistive-anode single-photon-counting photomultiplier, imaging spectrograph, Wollaston polarizer, multiparameter analyzer with histograming memory, and standard timing electronics. The spectrograph disperses the fluorescence spectrum across the photocathode of the photomultiplier, and the Wollaston polarizer separates the spectra of the two polarizations in opposite directions from the center of the photocathode perpendicular to the direction of spectral dispersion. The locations at which each photon reaches the photocathode is determined by the ratios of the charges read from the four corners of the resistive anode. One of the two address coordinates that determine where in histogramming memory each photon is recorded is obtained by measuring the time of arrival of the photon at the detector relative to the pulse of light that excites the fluorescence. The second address coordinate is obtained by combining the most-significant bit of the location of the event along the direction on the resistive anode corresponding to the polarization of the photon with the multibit digital value indicating photon wavelength. Storing the data directly into histogramming memory permits display of the data set as it is recorded. Both the spectral and temporal calibrations of the fluorescence analyzer are independent of the polarization of the fluorescence. The ≈100 ps temporal resolution of the resistive-anode detector is well matched to the ≈ 1 ns full width at half-maximum pulses of light produced by the synchrotron storage ring that we use as the excitation source, but laser excitation could also be used with this detector. Recording simultaneously all of the data required for the global analysis of the time evolution of both linear polarization components of fluorescence, and thus, time-resolved anisotropy, reduces the duration of exposure of the sample to the excitation beam, hence, facilitating studies of fragile or photosensitive biological specimens.

Publication Information

Output type

Scholary Output:
Contribution to journal
Article
Peer-review

Original language

English (US)

Pages from-to (Number of pages)

Pages 2279-2286 (8 pages)

Journal (Volume, Issue Number)

Review of Scientific Instruments (Volume 68, Issue 6)

Publication milestones

  • Published - 06/1997

Publication status

Published - 06/1997

ISSN

0034-6748

Publication IDs

  • Scopus: 0031167167

Publication metrics

Metrics

SciVal
citations
13
Scopus
citations
SciVal
FWCI
0.35
SciVal
Author count
3
SciVal
Paper percentile
64
Fractional count
1
Fractional count
0.33
Fractional count
2
Fractional count
0.67
Fractional count
1
Fractional count
1

PlumX, opens in new tab

Captures
7
Citation count
13