AES analysis of stainless steel corroded in saline, in serum and in vivo

David D. Zabel, Stanley A. Brown, Katharine Merritt, Joe H. Payer

Research output: Contribution to journalArticlepeer-review

28 Scopus citations

Abstract

Auger electron spectroscopic (AES) studies were conducted to examine the surface film of stainless steel specimens which had been subjected to passive and anodic corrosion in vitro and in vivo. Anodic corrosion was accelerated by the application of a +500 mV (SCE) potential for 30 minutes. Experiments performed in 0.9% saline indicated little alteration in the nature of the surface film compared to a freshly prepared control specimen. Auger analysis of specimens in 10% serum or in vivo revealed that passive film thickness was dependent on the corrosive environment. The films of specimens immersed under freely corroding conditions in 10% serum or in vivo were similar to the control film. Accelerated corrosion performed in 10% calf serum resulted in significant film thickening. The thickest films were from accelerated corrosion in vivo. The mechanism for the altered surface films produced by corrosion in vivo or in serum as compared to saline needs to be investigated further.

Original languageEnglish (US)
Pages (from-to)31-44
Number of pages14
JournalJournal of Biomedical Materials Research
Volume22
Issue number1
DOIs
StatePublished - Jan 1988
Externally publishedYes

ASJC Scopus subject areas

  • Biomaterials
  • Biomedical Engineering

Fingerprint

Dive into the research topics of 'AES analysis of stainless steel corroded in saline, in serum and in vivo'. Together they form a unique fingerprint.