TY - JOUR
T1 - Catheter Ablation of Ventricular Endocardium Using Radiofrequency Energy
T2 - Determinants of Lesion Volume and Shape
AU - GROGAN, E. WAYNE
AU - NELLIS, STEPHEN H.
AU - SUBRAMANIAN, RAMIAH
PY - 1989/8
Y1 - 1989/8
N2 - Radiofrequency energy was delivered at varying powers and durations to excised portions of canine left ventricle that were mounted in a tank of saline at room temperature. A radiofrequenciy generator delivered alternating current in a sine wave pattern at a frequency of 500 kHz, at variable voltage. In each of five excised ventricles, 49 lesions were created, at delivered RMS voltages from 18 to 40 V and durations of ablation from 5 to 180 seconds. Lesion diameter and depth were measured grossly, and lesion volume and shape (ratio of radius/depth) were calculated. Lesion diameter, depth, and volume increased with increasing power and duration of energy delivery, and the relation of delivered energy (joules) to calculated lesion volume (mm3) was linear (r = 0.88, P < 0.001). Shallower lesions were produced by shorter duration of energy delivery at all power levels, and similar volume but deeper, more symmetric lesions by lower power and longer duration of energy delivery. Radiofrequency energy in vitro thus produces reproducible lesions whose volume is proportional to delivered energy, and whose shape can be altered by delivering similar total energies, but varying the power and duration of energy delivery
AB - Radiofrequency energy was delivered at varying powers and durations to excised portions of canine left ventricle that were mounted in a tank of saline at room temperature. A radiofrequenciy generator delivered alternating current in a sine wave pattern at a frequency of 500 kHz, at variable voltage. In each of five excised ventricles, 49 lesions were created, at delivered RMS voltages from 18 to 40 V and durations of ablation from 5 to 180 seconds. Lesion diameter and depth were measured grossly, and lesion volume and shape (ratio of radius/depth) were calculated. Lesion diameter, depth, and volume increased with increasing power and duration of energy delivery, and the relation of delivered energy (joules) to calculated lesion volume (mm3) was linear (r = 0.88, P < 0.001). Shallower lesions were produced by shorter duration of energy delivery at all power levels, and similar volume but deeper, more symmetric lesions by lower power and longer duration of energy delivery. Radiofrequency energy in vitro thus produces reproducible lesions whose volume is proportional to delivered energy, and whose shape can be altered by delivering similar total energies, but varying the power and duration of energy delivery
KW - ablation
KW - arrhythmia
KW - catheter
KW - radiofrequency
UR - http://www.scopus.com/inward/record.url?scp=0024451188&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0024451188&partnerID=8YFLogxK
U2 - 10.1111/j.1540-8167.1989.tb01560.x
DO - 10.1111/j.1540-8167.1989.tb01560.x
M3 - Article
AN - SCOPUS:0024451188
SN - 1045-3873
VL - 3
SP - 243
EP - 251
JO - Journal of Cardiovascular Electrophysiology
JF - Journal of Cardiovascular Electrophysiology
IS - 4
ER -