TY - JOUR
T1 - Clinical experience with the Bjork-Shiley Delrin® tilting disc heart valve
AU - Dellsperger, Kevin C.
AU - Grunkemeier, Gary L.
AU - Stein, Paul D.
PY - 1996/8/1
Y1 - 1996/8/1
N2 - Background and aims of the study: The Bjork-Shiley Delrin® (BSD) disc heart valve was first used clinically in 1969. We estimate that as of January 1996, no more than 7,000 patients may be alive with the BSD valve. Methods: We reviewed the published reports of clinical experience with the BSD valve in addition to the records regarding BSD valves returned to Shiley Incorporated. Measurements of the maximum radial gap, static leak flow rate and inspection of the Delrin disc for wear were obtained. Results and conclusions: We concluded that clinically important regurgitation, due to disc wear, was rare but may occur. Larger prostheses, in general, showed greater disc wear than smaller sized valves. There were no reports of fracture of the Delrin disc. Two cases of inlet strut fracture were reported. Late disc embolization, in the absence of inlet strut fracture, occurred in two patients following surgical manipulation. Disc wear did not cause abnormal valve opening or closing although increased regurgitation may occur.
AB - Background and aims of the study: The Bjork-Shiley Delrin® (BSD) disc heart valve was first used clinically in 1969. We estimate that as of January 1996, no more than 7,000 patients may be alive with the BSD valve. Methods: We reviewed the published reports of clinical experience with the BSD valve in addition to the records regarding BSD valves returned to Shiley Incorporated. Measurements of the maximum radial gap, static leak flow rate and inspection of the Delrin disc for wear were obtained. Results and conclusions: We concluded that clinically important regurgitation, due to disc wear, was rare but may occur. Larger prostheses, in general, showed greater disc wear than smaller sized valves. There were no reports of fracture of the Delrin disc. Two cases of inlet strut fracture were reported. Late disc embolization, in the absence of inlet strut fracture, occurred in two patients following surgical manipulation. Disc wear did not cause abnormal valve opening or closing although increased regurgitation may occur.
UR - http://www.scopus.com/inward/record.url?scp=0029835424&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0029835424&partnerID=8YFLogxK
M3 - Article
C2 - 8905517
AN - SCOPUS:0029835424
SN - 0966-8519
VL - 5
SP - S169-S177
JO - Journal of Heart Valve Disease
JF - Journal of Heart Valve Disease
IS - SUPPL. 2
ER -