T1 - Comparison of time-of-flight mass spectrometry to triple quadrupole tandem mass spectrometry for quantitative bioanalysis
T2 - Application to antipsychotics
AU - Williamson, Leah N.
AU - Zhang, Guodong
AU - Terry, Alvin V.
AU - Bartlett, Michael G.
N1 - Funding Information:
This work was supported by the National Institute of Mental Health (MH066233). We would also like to thank Dr. Debra Gearhart and Dan Beck for providing the rat plasma samples.
PY - 2008/1
Y1 - 2008/1
N2 - Tandem mass spectrometry (MS/MS) is frequently used for the detection of small molecules in biological samples because of the high specificity and sensitivity associated with these instruments. Time-of-flight mass spectrometers (TOF-MS) are typically used qualitatively for accurate mass determination; however, recently they have gained more attention for quantitation. A selective and sensitive TOF-MS method was validated to determine risperidone, its active metabolite paliperidol, haloperidol, clozapine, and olanzapine in rat plasma using midazolam as an internal standard (IS). Comparisons were made between the use of LC-TOF and LC-MS/MS using a triple quadrupole for these compounds for specificity, linearity, precision, accuracy, matrix effects, and recovery. A focus of this study was the evaluation of recent instrumental and software improvements, which have been made to time-of-flight mass analyzers to increase the linearity and quantitative capabilities of these instruments.
AB - Tandem mass spectrometry (MS/MS) is frequently used for the detection of small molecules in biological samples because of the high specificity and sensitivity associated with these instruments. Time-of-flight mass spectrometers (TOF-MS) are typically used qualitatively for accurate mass determination; however, recently they have gained more attention for quantitation. A selective and sensitive TOF-MS method was validated to determine risperidone, its active metabolite paliperidol, haloperidol, clozapine, and olanzapine in rat plasma using midazolam as an internal standard (IS). Comparisons were made between the use of LC-TOF and LC-MS/MS using a triple quadrupole for these compounds for specificity, linearity, precision, accuracy, matrix effects, and recovery. A focus of this study was the evaluation of recent instrumental and software improvements, which have been made to time-of-flight mass analyzers to increase the linearity and quantitative capabilities of these instruments.
KW - Antipsychotics
KW - LC-MS
KW - Method validation
KW - Time-of-flight mass spectrometry
UR - http://www.scopus.com/inward/record.url?scp=52449125833&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=52449125833&partnerID=8YFLogxK
U2 - 10.1080/10826070802388243
DO - 10.1080/10826070802388243
M3 - Article
AN - SCOPUS:52449125833
SN - 1082-6076
VL - 31
SP - 2737
EP - 2751
JO - Journal of Liquid Chromatography and Related Technologies
JF - Journal of Liquid Chromatography and Related Technologies
IS - 18
ER -