Abstract
There is significant interest in delivering precisely targeted small-volume radiation treatments, in the pre-clinical setting, to study dose-volume relationships with tumor control and normal tissue damage. In this work we investigate the IGRT targeting accuracy of the XRad225Cx system from Precision x-Ray using high resolution 3D dosimetry techniques. Initial results revealed a significant targeting error of about 2.4mm. This error was reduced to within 0.5mm after the IGRT hardware and software had been recalibrated. The facility for 3D dosimetry was essential to gain a comprehensive understanding of the targeting error in 3D.
Original language | English (US) |
---|---|
Article number | 012070 |
Journal | Journal of Physics: Conference Series |
Volume | 444 |
Issue number | 1 |
DOIs | |
State | Published - 2013 |
Externally published | Yes |
Event | 7th International Conference on 3D Radiation Dosimetry, IC3DDose 2012 - Sydney, NSW, Australia Duration: Dec 4 2012 → Dec 8 2012 |
ASJC Scopus subject areas
- General Physics and Astronomy